Europa Range

Europa Range

Europa Range

High voltage rack mounted power supply

Laboratory power supply – 20, 35, 50 & 100kV

Description

The Europa range of power supplies are a classic high voltage power supply design for laboratory applications. Up to 100kV output voltage is available and both positive and negative polarities can be provided. Extremely accurate output parameters are achieved as can be seen from the specification. The power supply is also well protected against erratic line and load conditions. The reversible section is removable in the form of a drawer accessible from the front of the unit. Two sections – one positive and one negative are exchanged to engage the alternative polarity. Control and monitoring can be via an IEEE interface or by front panel potentiometers.

 

 

Performance

V I W
100kV 1mA 100W
50kV 4mA 200W
35kV 6mA 210W
20kV 10mA 200W
Line regulation Not more than 50ppm of maximum rated output voltage for
± 10% input line change
Load regulation Not more than 100ppm of maximum rated output voltage
for 0 to maximum output current change.
Ripple Not more than 20V (peak to peak) – 100W Version

Temperature coefficient

(0 to 50°C) Not more than 50ppm of maximum output per °C
Stability after 1/2 hour warm up Not more than 0.01% per hour
Not more than 0.02% per 8 hours.

Front Panel Feature

Front panel meters Front panel meters are available for reading output voltage
and current with accuracy of ± 2% of full scale.
Front panel controls Front panel controls are available for output voltage and current control.
Protection The power supply is protected against overloads, short circuits and arcs.

Mechanical Features

Size 19 inch wide suitable for standard rack mounting, 6U high
Weight Main Unit: 35kg
Alternative Polarity Head: 12kg
High voltage connector and cable Pre-assembled mating high voltage connector with 3 metres
of high voltage shielded cable will be provided as standard.
Power input connector A 2m 3-wire mains lead plug will be provided as standard.
Cooling Forced air by means of a fan.
IEEE-488 interface Full talker/listener capability is available through the IEEE-488 interface.
The interface functions allow output voltage and current limit to be programmed via the IEEE-488 bus. In addition, voltage and current measurements can be taken on request.
Power requirements 230V AC ± 10%, less than 2A, 50Hz.

Enviromental Requirements

Temperature Operating: 15°C to 35°C.
Storage: 0°C to 50°C.
Relative humidity Operating: 30% to 80%.
Storage: < 95%.